The biological evaluation and testing requirements, combined with the EO-sterilisation process residue limits, form the justification that an EO-sterilised device is acceptable for use.
PRINCIPLE OF TEST:a) Limited exposure – devices whose cumulative single, multiple or repeated use or contact is up to 24h.
b) Prolonged exposure – devices whose cumulative single, multiple or repeated long-term use or contact
is likely to exceed 24h but not 30d.
c) Permanent contact – devices whose cumulative single, multiple or repeated use or contact exceeds 30d.
The guiding principle in selecting appropriate extraction methods for the determination of EO is the evaluation of the dose to the patient in order to show compliance with the requirements set out in ISO 10993-7:2008, using simulated use wherever possible. For devices in the prolonged exposure category, it is important to note that the device must also meet the residue requirements of the limited exposure category, and that devices in the permanent contact category must also meet the residue requirements of the prolonged exposure and limited exposure categories, whichever extraction condition is used. Where residues are shown to be within these requirements for products tested by exhaustive extraction, there is no need to further challenge the device by simulated-use extraction.
SAMPLE REQUIREMENTS:
Three blank samples (unsterilised) of each product are required for controls and validation work. Client decides how many sterilised samples for each product to test. The more replicate of results, the more analytically sound the result will be. As a minimum, two replicates for each product from each aeration time are recommended.
TURNAROUND TIMES:
Five working days from receipt of samples and signed submission form given no further aeration time is needed.
